LND 712
END WINDOW-ALPHA-BETA-GAMMA DETECTOR

GENERAL SPECIFICATIONS

Gas Filling Ne +Halogen
Cathode Material 446 Stainless Steel
Maximum Length (inch/mm) 1.94 / 49.2
Effective Length (inch/mm) 1.5 / 38.1
Maximum Diameter (inch/mm) 0.59 / 15.1
Effective Diameter (inch/mm) 0.36 / 9.1
Connector Pin
Operating Temperature Range 0C -40 to +75

WINDOW SPECIFICATIONS

Areal Density (mg/cm2) 1.5 - 2.0
Effective Diameter (inch/mm) 0.36 / 9.1
Material Mica

ELECTRICAL SPECIFICATIONS

Recommended Anode Resistor (meg ohm)   circuit diagram 10
Maximum Starting Voltage (volts) 325
Recommended Operating Voltage (volts) 500
Operating Voltage Range (volts) 450 - 650
Maximum Plateau Slope (%/100 volts) 6
Minimum Dead Time (micro sec) 90
Gamma Sensitivity Co60 (cps/mR/hr) 18
Maximum Background Shielded 50mmPb + 3mmAl (cpm) 10
Tube Capacitance (pf) 3
Weight (grams) 8